Sigma SQC-122c manual Appendix O Connections, Connector Wiring Relay Pins, Input Pins

Page 61

Appendix

C.I/O Connections

A 25 pin D-sub connector, located on the rear of the SQC-122c, provides Input/Output connections. Refer to Section 3.6 for relay and input functional assignments.

Inputs can be activated either by connecting to a switch and shorting to Ground, or they can be driven by a TTL compatible signal.

WARNING: These are not isolated inputs! The voltage level applied must be limited to between 0 and +5 volts with respect to Ground.

WARNING: Output relays are rated for 30Vrms or 30VDC, at 2A maximum. Proper fusing, and adequate wiring insulation and separation, should be provided to assure these limits are not exceeded.

The pin assignments for the rear panel mounted I/O connector are shown below:

1

13

14

25

I/O Connector Wiring

Relay

Pins

Relay 1

14,15

Relay 2

1,2

Relay 3

3,4

Relay 4

5,6

Relay 5

7,8

Relay 6

9,10

Relay 7

11,12

Relay 8

13,25

 

 

Input

Pins

Input 1

16

Input 2

17

Input 3

18

Input 4

19

Input 5

20

Input 6

21

Input 7

22

Input 8

23

Ground

24

Image 61
Contents SQC-122c Thin Film Deposition Controller User’s Guide Safety Symbols Safety InformationLimitation of Warranty Warranty InformationReturn Policy Chapter Operation Table of Contents Chapter Quick StartChapter Menus Chapter OptionsAppendix Thin Film Process Overview Chapter Quick Start IntroductionChapter Quick Start Front Panel Chapter Quick Start Rear Panel Power Input FuseChapter Quick Start System Connections Chapter Quick Start Installation Power Chapter Quick Start Process SetupQuick Setup SelectChapter Quick Start Edit Mode LayerExit to Main Chapter Quick Start Depositing a Film This page left blank for your notes Definitions Chapter Operation IntroductionDefining a Film POWER% POWER% Deposition Controls Menu Layer Edit Menu Chapter Operation Defining a ProcessChapter Operation Chapter Operation Sensor Setup Tooling U n d e r 1 0 0 % Chapter Operation Chapter Operation Source Setup Chapter Operation Running a Process Chapter Operation Next Menu Zero THICKNESSkA RATEA/s 824 55.3 Chapter Operation Loop Tuning Chapter Operation No Readings, or Erratic Readings from Sensors Chapter Operation TroubleshootingPoor Rate Stability Incorrect Rate or Thickness MeasurementThis page left blank for your notes Chapter Menus Introduction Chapter Menus Main Menu Default, power-up Main Screen display is shown belowNext SetupMenu Quick FilmChapter Menus Next Menu Screens Exit to Main Edit Chapter Menus Quick Setup MenuPrev Layer Next LayerChapter Menus Main Screen Prev Menu Select Edit Delete Control Knob Chapter Menus Process MenusInsert Layer Control Knob Main Screen Prev Menu Edit Cut PasteExit to Main Prev Menu Edit Control Knob POWER%Chapter Menus Main Screen Prev Menu Edit Delete Control Knob Chapter Menus Film MenusExit to Main Prev Menu Edit Exit to Main Prev Menu Edit THICKNESSkA RATEA/s Deposition Controls Menu Chapter Menus Chapter Menus Chapter Menus Chapter Menus System Parameters Menu Chapter Menus Chapter Menus Mapping Menu Chapter Menus I/O SetupInputs Explanation high input causes Relays Explanation closes whenThis page left blank for your notes Chapter Options Introduction Full Rack Extender InstallationOption Card Appendix Material Parameters General Specifications Appendix SpecificationsMeasurement SourceDisplay Input Pins Connector Wiring Relay PinsAppendix O Connections Appendix Appendix Handheld Remote Controller Appendix EC Declaration of Conformity Complies with the Essential Health and Safety RequirementsDeclares that the product Uses the following standards