DIAGnostic:COMMunicate :SERial[0][:OWNer]?

:COMMunicate

DIAGnostic:COMMunicate:SERial[ 0] [ :OWNer] ? Returns the current "owner"

:SERial[0][:OWNer]?

of the built-in serial interface. The values returned will be; "SYST", "IBAS", or

 

"NONE".

 

Comments

Related Commands: DIAGnostic:SERial[:OWNer]

Example Determine which instrument has the serial interface.

 

 

DIAG:COMM:SER?

Note; :OWNer is implied

 

 

enter statement

statement returns the string

 

 

 

SYST, IBAS, or NONE

 

 

:COMMunicate DIAGnostic:COMMunicate:SERial[n]:STORe Stores the serial

:SERial[n]:STORe

communications parameters (e.g. BAUD, BITS, PARity etc.) into non-volatile

 

storage for the serial interface specified by [n] in SERial[n].

Comments

Until … STORe is executed, communication parameter values are stored

 

 

in volatile memory, and a power failure will cause the settings to be lost.

 

DIAG:COMM:SER(1-7):STOR causes an Agilent E1324A (B-size

 

 

RS-232 card) to store its settings in an on-board EEROM. This EEROM

 

 

write cycle takes nearly one second to complete. Wait for this operation

 

 

to complete before attempting to use that serial interface.

 

• The Agilent E1324A’s EEROM used to store its serial communication

 

 

settings has a finite lifetime of approximately ten thousand write cycles.

 

 

Even if your application program sent the … STORe command once

 

 

every day, the lifetime of the EEROM would still be over 27 years.

 

 

Be careful that your application program sends the … STORe command

 

 

to an Agilent E1324A no more often than is necessary.

 

Related Commands: all SYST:COMM:SER[n]… commands

Example Store the serial communications settings in the third Agilent E1324A.

DIAG:COMM:SER3:STOR

7-8 System Instrument Command Reference