In Case of Difficulty

Problem:

Problem Cause/Solution:

 

 

Error -113 undefined header error occurs after entering

For some commands used by the menus, the data

data in response to a menu prompt.

entered is appended to a command header. For

 

example, if you enter "1" as the port number for a digital

 

I/O module, the command used is

 

DIG:HAND1:MODE NONE where HAND1 indicates

 

the port number. If your entry was invalid or incorrect,

 

error -113 occurs.

 

 

Following the power-on sequence or system reset the

An unnassigned device (incorrect logical address) was

display shows:

detected, or the contents of non-volatile memory may

 

have been lost, If you cycle power or perform system

Configuration errors. Select SYSTEM

reset, the display will show the logical address of the

Press any key to continue_

unassigned device. You can also check the logical

 

addresses using the CONFIG? -- LADDS branch of the

 

System Instrument menu. Refer to Chapter 1 of this

 

manual for a discussion of logical addresses and

 

unassigned devices.

 

 

The display shows: "instrument in local lockout".

The front panel has been locked-out (GPIB local

Menus seem to work but nothing happens when I reach

lockout). You can re-enable menu operation by

the bottom level or try to execute a command.

cancelling local lockout (from remote) or by cycling

 

mainframe power.

 

 

Display cannot be removed from monitor mode.

Monitor mode was entered from remote

 

(DISP:MON:STAT ON command) and the front panel

 

has also been locked out (GPIB local lockout). Either

 

cancel the local lockout or execute

 

DISP:MON:STAT OFF (from remote).

 

 

Display shows:

A hardware or software problem has occured in the

 

instrument preventing it from responding to front panel

Can not connect to instrument

control.

Press any key to continue_

 

 

 

After selecting an instrument the display shows:

The instrument is busy performing an operation. Press

 

Clear Instr to abort the instrument operations and

busy.

allow the front panel to access the instrument.

 

 

Display shows:

The instrument has already been selected from the

 

Display Terminal Interface. An instrument can only be

Instrument in use by another display.

“attached” to one display at a time. At the terminal,

Press any key to continue_

return to the “Select instrument” menu. The instrument

 

can now be selected from the Front Panel.

 

 

 

 

2-12 Using the Front Panel

Page 32
Image 32