*IDN? Identity. Returns the device identity. The response consists of the following four fields (fields are separated by commas):

Manufacturer

Model Number

Serial Number (returns 0 if not available)

Firmware Revision (returns 0 if not available)

 

The *IDN? command returns the following command string for the E1301B:

 

AGILENT,E1301B,0,A,07.00

 

This command will return the following string for the E1300B:

 

AGILENT,E1300B,0,A,07.00

NOTE

 

The revision will vary with the revision of the ROM installed in the system. This

 

is the only indication of which version of ROM is in the box. The major number

 

(01 in the examples) indicates whether there have been functional changes made

 

in this ROM. The minor number (00 in the examples) indicates whether only

 

bug fixes and minor changes were made.

 

 

Example Get the ID fields from the system and print them.

10 DIM A$[50]

Dimension array for ID fields

20 OUTPUT 70900;"*IDN?"

Queries identity

30 ENTER 70900;A$

Places ID fields in array

40 PRINT A$

Print ID fields

50END

*LMC? Learn Macros Query. Returns a quoted string name for each currently defined macro. If more than one macro is defined, the quoted strings are separated by commas (,). If no macro is defined, then a quoted null string ("") is returned.

*LRN? Learn query command. *LRN? causes the instrument to respond with a string of SCPI commands which define the instrument’s current state. Your application program can enter the *LRN? response data into a string variable, later to be sent back to the instrument to restore that configuration.

Example response from an Agilent E1326B voltmeter in the power-on state:

*RST;:CAL:ZERO:AUTO 1; :CAL:LFR + 60; VAL + 0.00000000E+ 000; :DISP:MON:STAT 0; CHAN (@0); :FORM ASC,+ 7; :FUNC "VOLT"; :MEM:VME:ADDR + 2097152; SIZE + 0; STAT 0; :RES:APER

+1.666667E -002; OCOM 0; RANG + 1.638400E+ 004; RANG:AUTO 1;:VOLT:APER + 1.666667E -002; RANG + 8.000000E+ 000; RANG:AUTO 1; :TRIG:COUN + 1; DEL + 0.00000000E+ 000; DEL:AUTO 1; :TRIG:SOUR IMM; :SAMP:COUN + 1; SOUR IMM;TIM + 5.000000E -002 S