Table B-2. Error Messages and Causes

Error Messages and Causes

Code

Message

Cause

 

 

 

-101

Invalid character

Unrecognized character in specified parameter.

- 102

Syntax error

Command is missing a space or comma between

 

 

parameters

- 103

Invalid separator

Command parameter is separated by some

 

 

character other than a comma.

- 104

Data type error

The wrong data type (i.e. number, character,

 

 

string expression) was used when specifying a

 

 

parameter.

- 108

Parameter not allowed

Parameter specified in a command which does

 

 

not require one.

- 109

Missing parameter

No parameter specified in the command in

 

 

which a parameter is required.

- 113

Undefined header

Command header was incorrectly specified.

- 123

Numeric overflow

A parameter specifies a value greater than the

 

 

command allows.

- 128

Numeric data not allowed

A number was specified for a parameter when a

 

 

letter is required.

- 131

Invalid suffix

Parameter suffix incorrectly specified

 

 

(e.g. .5SECOND rather than .5S or .5SEC).

- 138

Suffix not allowed

Parameter suffix is specified when one is not

 

 

allowed.

- 141

Invalid character data

The discrete parameter specified is not allowed

 

 

(e.g. TRIG:SOUR INT - INT is not a choice.)

- 178

Expression data not allowed

A parameter other than the channel list is

 

 

enclosed in parentheses.

- 211

Trigger ignored

Trigger occurred while the Pacer is in the idle

 

 

state, or a trigger occurred from a source other

 

 

than the specified source.

- 222

Data out of range

The parameter value specified is too large or too

 

 

small.

- 224

Illegal parameter value

The numeric value specified is not allowed.

- 240

Hardware error

Hardware error detected during power-on cycle.

 

 

Return multimeter to Agilent for repair.

- 310

System error

If caused by *DMC, then macro memory is full.

- 350

Too many errors

The error queue is full as more than 30 errors

 

 

have occured.

- 410

Query interrupted

Data is not read from the output buffer before

 

 

another command is executed.

- 420

Query unterminated

Command which generates data not able to

 

 

finish executing due to a multimeter

 

 

configuration erorr.

- 430

Query deadlocked

Command execution cannot continue since the

 

 

mainframe’s command input, and data output

 

 

buffers are full. Clearing the instrument restores

 

 

control.

1500

External trigger source

"Event In" signal already allocated to another

 

already allocated

instrument such as a Switchbox.

2002

Invalid logical address

A value less than 0 or greater than 255 was

 

 

specified for logical address.

2003

Invalid word address

An odd address was specified for a 16 bit read or

 

 

write. Always use even addresses for 16 bit

 

 

(word) accesses.

2005

No card at logical address

A non-existent logical address was specified with

 

 

the VXI:READ? or VXI:WRITE command.

2101

Failed Device

VXI device failed its self test.

2102

Unable to combine device

Device type can not be combined into an

 

 

instrument such as a scanning voltmeter or a

 

 

switchbox.

2103

Config warning, Device

ID of device does not match list of drivers

 

driver not found

available. Warning only.

2105

Config error 5, A24 memory

More A24 memory installed in the mainframe

 

overflow

than can be configured into the available A24

 

 

memory space.

2108

Config error 8, Inaccessible

A24 memory device overlaps memory space

 

A24 memory

reserved by the mainframe’s operating system.

Error Messages B-3