Error Messages and Causes

Code

Message

Cause

 

 

 

2110

Config error 10, Insufficient

Too many instruments installed for the amount

 

system memory

of RAM installed in the mainframe. Cannot

 

 

configure instruments. Only the system

 

 

instrument is started.

2111

Config error 11, Invalid

A device’s logical address is not a multiple of 8

 

instrument address

and the device is not part of a combined

 

 

instrument.

2113

Config error 13, Logical

Duplicate logical addresses set or interrupt

 

address or IACK switch set

bypass switches set improperly. Only the system

 

wrong

instrument is started.

2129

Config warning, Sysfail

A device was asserting SYSFAIL on the

 

detected

backplane during startup.

2130

Config error 30, Pseudo

A physical device has the same logical address as

 

instrument logical address

IBASIC (240)

 

unavailable

 

2131

Config error 32, File system

Insufficient system resources to allow the

 

start up failed

IBASIC file system to start.

2145

Config warning, Non-volatile

NVRAM was corrupted or a cold boot was

 

RAM contents lost

executed.

2148

Config warning, Driver RAM

Driver RAM was corrupted or a cold boot was

 

contents lost

executed.

2202

Unexpected interrupt from

A register based card interrupted when an

 

non-message based card

interrupt service routine had not been set up.

2809

Interrupt line has not been

A DIAG:INT:ACT or DIAG:INT:RESP

 

set up

command was executed before setting the

 

 

interrupt with DIAG:INT:SET.

 

 

 

B-4 Error Messages