INITiate [:IMMediate]

INITiate

The INITiate command subsystem controls the initiation of the trigger system

 

for one or more trigger cycles. INITiate enables while ABORt disables the

 

trigger system. The TRIGger command subsystem controls the behavior of the

 

trigger system while it is enabled.

 

Subsystem Syntax

INITiate

 

 

[:IMMediate]

 

 

 

[:IMMediate] INITiate:IMMediate changes the trigger system from the Idle state to the Wait

 

For Trigger state.

 

Comments

• If TRIGger:SOURce is IMMediate, the Pacer starts. If TRIG:SOURce is

 

BUS, EXT, or HOLD, the Pacer will start when that trigger condition is

 

satisfied.

 

 

• Sending the ABORt command will reset the trigger system back to its

 

Idle state and terminate any pacer pulse train in progress.

 

• Sending INIT while the system is still in the Wait for Trigger state

 

(already INITiated) will cause an error -213,"Init ignored".

 

Related Commands: ABORt, TRIGger

 

 

*RST Condition: Trigger system is in the Idle state.

Example Initiating the trigger system (Wait For Trigger state).

 

TRIG:SOUR HOLD

trigger source is TRIG

 

 

command

 

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

 

SOUR:PULS:PER .1 S

pulse period set to .1 second

 

INIT

go to Wait For Trigger state

 

TRIG

trigger the Pacer to output

 

 

pulses

 

.

 

 

.

 

 

INIT

must re-initiate system before

each trigger cycle

TRIG

.

.

System Instrument Command Reference 7-29

Page 157
Image 157