Agilent Technologies 75000 SERIES B, E1301B, E1300B TRIGger DELay, Trigdelay .75 S

Models: E1300B E1301B 75000 SERIES B

1 244
Download 244 pages 1.94 Kb
Page 179
Image 179

 

 

 

 

 

 

TRIGger :DELay

TRIGger

 

The TRIGger command subsystem controls the behavior of the trigger system

 

 

once it is initiated (see INITiate command subsystem). The trigger command

 

 

subsystem controls:

 

 

 

 

 

 

• The delay between trigger and first Pacer pulse (TRIG:DELay)

 

 

 

 

• An immediate software trigger (TRIG:IMM)

 

 

 

 

The source of the trigger (TRIG:SOUR BUS EXT HOLD IMM)

Subsystem Syntax

 

TRIGger

 

 

 

 

 

 

:DELay < delay>

 

 

 

 

 

:DELay? [MIN MAX]

 

 

 

 

 

[:IMMediate]

 

 

 

 

 

 

:SLOPe < slope>

 

 

 

 

 

:SLOPe?

 

 

 

 

 

 

:SOURce BUS EXT HOLD IMM

 

 

 

 

:SOURce?

 

 

 

 

:DELay

 

 

 

 

 

 

 

TRIGger:DELay < delay> sets the delay between receipt of trigger and first

 

 

Pacer pulse.

 

 

 

 

Parameters

 

 

 

 

 

 

 

 

Parameter

 

Parameter

Range of

Default

 

 

 

 

Name

 

Type

Values

Units

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

delay

 

numeric

250E -9s to 4.19430375s or

second

 

 

 

 

 

 

 

MIN MAX

 

 

Comments

 

The resolution for delay is 250E -9 seconds.

 

 

 

 

Related Commands: ABORt, INITiate

 

 

 

 

*RST Condition: TRIG:DELay 2.5E -9

 

 

Example Setting delay between trigger and Pacer output.

 

 

 

 

 

TRIG:SOUR HOLD

trigger is TRIG command

 

 

 

SOUR:PULS:COUN 100

set Pacer to output 100 pulses

 

 

 

SOUR:PULS:PER .1 S

pulse period set to .1 second

 

 

 

TRIG:DELAY .75 S

start Pacer .75 sec after trigger

 

 

 

INIT

 

go to Wait For Trigger state

 

 

 

TRIG

 

trigger Pacer to output pulses

:DELay? TRIGger:DELay? [MIN MAX] returns:

The current delay if no parameter is sent.

The maximum allowable delay if MAX is sent.

The minimum allowable delay if MIN is sent.

Example Querying the trigger delay setting.

 

TRIG:DEL .75 S

start Pacer .75 sec after trigger

TRIG:DEL?

command System Instrument

 

to send TRIG:DEL value.

enter statement

input value of trigger delay

System Instrument Command Reference 7-51

Page 179
Image 179
Agilent Technologies 75000 SERIES B, E1301B, E1300B user manual TRIGger DELay, Trigdelay .75 S