:DOWNload:CHECked

:SADDress

Parameters

DIAGnostic :DOWNload:CHECked :SADDress

DIAGnostic:CHECked:DOWNload:SADDress < address> ,< data> writes data to non-volatile User RAM at a single address specified by address using error correction. It can also write to devices with registers in the A16 address space.

Parameter

Parameter

Range of

Default

Name

Type

Values

Units

 

 

 

 

address

numeric

0 to 16,777,215 (# HFFFFFE)

none

 

 

 

 

data

arbitrary block

See "Parameter Types", in the

none

 

program data

beginning of this chapter

 

Comments

• This command is typically used to send data to a device which accepts

 

data at a single address. It is the only way to send binary data to single

 

addresses over a serial (RS232C) line.

 

• Most computers terminate an OUTPUT, PRINT, or WRITE statement

 

with a carriage return or carriage return and line feed. These

 

End-Of-Line characters must be either accounted for (NRAM segment

 

sized to accommodate them), or suppressed using an appropriate

 

IMAGE or FORMAT statement. Some helpful methods:

 

Control the End-Of-Line characters with format statements.

 

Use the Definite Length Arbitrary Block Program Data format (see

 

example) to send your data rather than the Indefinite Length

 

Arbitrary Block Program Data format.

 

• A register address in A16 address space can be determined by:

 

1FC0016 + (LADDR * 64) + register_number

 

where 1FC00016 is the base address in the System Instrument A16 space,

 

LADDR is the device logical address, 64 is the number of address bytes

 

per device, and register_number is the register to which the data is

 

written.

If the device is an A24 device, the address can be determined using the VXI:CONF:DLISt command to find the base address in A24, and then adding the register_number to that value. A24 memory between address 20000016 and address E0000016 is directly addressable by the Controller.

Address may be specified in decimal, hex (# H), octal (# Q), or binary (# B) formats. DOWNload is done by word (16 bit) access so address must be even.

Related Commands: DIAG:UPLoad:SADDress?

System Instrument Command Reference 7-11