Start-up Error Messages

Start-up errors are most often generated just after the mainframe is powered-up or re-booted (DIAG:BOOT command). If you have an Agilent E1301B, or an Agilent E1300B with a terminal connected to the Display Terminal Interface (built-in RS-232 only), you can read these errors on the front panel or terminal. If you have an Agilent E1300B and no terminal, then you must access this error information by sending the VXI:CONF:DLIS? command over GPIB. We recommend that users of either model include a routine at the beginning if their application program which checks for start-up errors before the program trys to access individual instruments. See your Installation and Getting Started Guide for an example program.

Table B-3. Start-up Error Messages and Warnings

Start-Up Error Messages and Warnings

Code

Message

Cause

 

 

 

 

 

1

Failed Device

VXI device failed its self test.

 

2

Unable to combine device

Device type can not be combined into an

 

 

 

instrument such as a scanning voltmeter or a

 

 

 

switchbox.

 

3

Config warning, Device

ID of device does not match list of drivers

 

 

driver not found

available. Warning only.

 

5

Config error 5, A24 memory

More A24 memory installed in the mainframe

 

 

overflow

than can be configured into the available A24

 

 

 

memory space.

 

8

Config error 8, Inaccessible

An A24 memory device overlaps a memory

 

 

A24 memory

space reserved by the mainframe’s operating

 

 

 

system.

 

10

Config error 10, Insufficient

Too many instruments installed for the amount

 

 

system memory

of RAM installed in the mainframe. Cannot

 

 

 

configure instruments. Only the system

 

 

 

instrument is started.

 

11

Config error 11, Invalid

A device’s logical address is not a multiple of 8

 

 

instrument address

and the device is not part of a combined

 

 

 

instrument.

 

13

Config error 13, Logical

Duplicate logical addresses set or interrupt

 

 

address or IACK switch set

bypass switches set improperly. Only the system

 

 

wrong

instrument is started.

 

29

Config warning, Sysfail

A device was asserting SYSFAIL on the

 

 

detected

backplane during startup.

 

30

Config error 30, Pseudo

A physical device has the same logical address as

 

 

instrument logical address

IBASIC (240)

 

 

unavailable

 

 

31

Config error 32, File system

Insufficient system resources to allow the

 

 

start up failed

IBASIC file system to start.

 

45

Config warning, Non-volatile

NVRAM was corrupted or a cold boot was

 

 

RAM contents lost

executed.

 

48

Config warning, Driver RAM

Driver RAM was corrupted or a cold boot was

 

 

contents lost

executed.

 

 

 

 

 

Error Messages B-5