VXI :CONFigure:DLISt?

:CONFigure:DLISt? VXI:CONF:DLIS? [< logical_addr> ] returns information about the device specified by logical_addr. Response data is in the form:

n1, n2, n3, n4, n5, n6, c1, c2, c3, c4, c5, s1, s2, s3, s4

Where the fields above are defined as:

n fields

Indicate numeric data response fields.

c fields

Indicate character data response fields.

s fields

Indicate string data response fields.

n1 Device’s Logical Address. A number from 0 to 255.

n2 Commander’s Logical Address. A number from -1 to 255; -1 means this device has no commander.

n3 Manufacturer’s ID. A number from 0 to 4095.

n4 Model Code. A number from 0 to 65535, chosen by the manufacturer to signify the model of this device.

n5 Slot Number. A number between -1 and the number of slots in this mainframe; -1 indicates that the slot associated with this device is unknown. This is always -1 for B size mainframes.

n6 Slot 0 Logical Address. A number from 0 to 255.

c1 Device Class. 3 data characters; EXT HYB MEM MSG REG VME . EXT = Extended device, HYB = hybrid device (e.g. IBASIC), MEM = memory device, MSG = Message-based device,

REG = Register-based device, VME = VME device

c2 Memory Space. Up to 4 data characters; A16 A24 A32 NONE RES. A16 = A16 addressing mode, A24 = A24 addressing mode, A32 =

A32 addressing mode, NONE = no addressing mode, RES = reserved.

c3 Memory Offset. 10 data characters which define the base address of the A24 or A32 address space on the device. This value is expressed in hex format (first two characters are # H).

c4 Memory Size. 10 data characters which define the size of the A24 or A32 address space in bytes. This value is expressed in hex format (first two characters are # H).

c5 Pass/Failed. Up to 5 data characters which define the status of the device; FAIL IFAIL PASS READY. FAIL = failed self-test, IFAIL = configuration register initialization fails,

PASS = self-test passed, READY = ready to receive commands

s1 Extended Field 1. Not currently used; returns ""

s2 Extended Field 2. Not currently used; returns ""

s3 Extended Field 3. Not currently used; returns ""

s4 Manufacturer’s Specific Comments. Up to 80 character string contains manufacturer specific data in string response data format. This field is sent with a 488.2 string response data format, and will contain the instrument name and its IEEE 488.1 secondary address unless a start-up error is detected. In that case, this field will contain one or more error codes in the form "CNFG ERROR: n, m, ...,z" . See Appendix B, Table B-3 for a complete list of these codes.

System Instrument Command Reference 7-55

Page 183
Image 183
Agilent Technologies E1301B, E1300B VXI CONFigureDLISt?, N1, n2, n3, n4, n5, n6, c1, c2, c3, c4, c5, s1, s2, s3, s4