DIAGnostic :UPLoad[:MADDress]?

:UPLoad[:MADDress]?

DIAGnostic:UPLoad[:MADDress]? < address> ,< byte_count> Returns the

 

number of bytes specified by byte_count, starting at address.

 

Parameters

 

 

 

 

 

Parameter

 

Parameter

Range of

Default

 

Name

 

Type

Values

Units

 

 

 

 

 

 

 

address

 

numeric

0 to 16,777,215 (# HFFFFFE)

none

 

 

 

 

 

 

 

byte_count

 

numeric

0 to (999,999,998)

none

Comments

A ddress may be specified in decimal, hex (# H), octal (# Q), or binary

 

(# B) formats.

 

 

 

UPLoad is done by word (16 bit) access so address and byte_count must be even.

Data is returned in the Definite Block Response Data format:

# < non-zero digit> < digit(s)> < data byte(s)>

Where the value of < non-zero digit> equals the number of < digit(s)> . The value of < digit(s)> taken as a decimal integer indicates the number of < data byte(s)> to expect in the block.

This command can also be used to retrieve data from a device with registers in A16 address space. See DIAG:UPload:SADDress?

Related Commands: DIAG:NRAM:ADDress?, DIAG:NRAM:CREate, DIAG:DOWNload

Example Upload data stored on non-volatile User RAM.

DIM HEADER$[6],DATA(1024)

6 chars for "# 41024" header

1024 chars for data bytes

DIAG:NRAM:ADDR?

get starting address of NRAM

enter ADD

address into ADD

OUTPUT "DIAG:UPL? < value of ADD> ,1024"

request 1 Kbyte from address in ADD

enter HEADER$

strip "# 41024" from data

enter DATA

get 1024 data bytes into string; use enter format so statement won’t terminate on CRs or LFs etc. Line Feed (LF) and EOI follow the last character retrieved.