*RCL < state number> Recall stored state. Recalls a stored state from memory and configures the instrument to that state. States are stored using the *SAV command.

Example

OUTPUT 70900;"*RCL 4"

Recalls instrument state

 

 

number 4

*RMC < name_string> Remove Individual Macro Command. Purges an individual macro identified by the name_string parameter.

Example

output 70900;"*RMC ’LIST’"

remove macro command from

 

 

*DMC example

NOTE: At printing time, *RMC is a command proposed for a revision and re-designation of ANSI/IEEE Std 488.2-1987.

*RST Reset. Resets an instrument as follows:

Sets the instrument to a known state (usually the power-on state)

Aborts all pending operations

Disables the *OPC and *OPC? modes.

*RST does not affect:

The state of the GPIB interface

The GPIB address

The output queue

The Service Request Enable Register

The Standard Event Status Enable Register

The power-on flag

Calibration data

Protected user data

*SAV < state number> Store state. Stores an instrument’s present state in a numbered memory location (< state number> parameter). State numbers can range from 0 to 9.

Example

OUTPUT 70900;"*SAV 4"

Saves present instrument state

 

 

as state number 4

*SRE < mask> Service Request Enable. When a service request event occurs, it sets a corresponding bit in the Status Byte Register (this happens whether or not the event has been enabled (unmasked) by *SRE). The *SRE command allows you to identify which of these events will assert an GPIB service request (SRQ). When an event is enabled by *SRE and that event occurs, it sets a bit in the Status Byte Register and issues an SRQ to the computer (sets the GPIB SRQ line true). You enable an event by specifying its decimal weight for < mask> . To enable more than one event, specify the sum of the decimal weights. Refer to "The Status Byte Register" earlier in this chapter for a table showing the contents of the Status Byte Register.

Example

OUTPUT 70900;"*SRE 160"

Enables bits 5 & 7. Respective

 

 

weights are 32 + 128 = 160