Agilent Technologies E1300B, E1301B user manual TRIGger SOURce?, Parameter Value Source of Trigger

Models: E1300B E1301B 75000 SERIES B

1 244
Download 244 pages 1.94 Kb
Page 181
Image 181

 

 

 

TRIGger :SOURce?

Comments

• The following table explains the possible choices.

 

 

 

 

 

 

Parameter Value

Source of Trigger

 

 

 

 

 

 

BUS

Group Execute Trigger (GET) bus command,

 

 

 

*TRG common command, or TRIGger

 

 

 

command.

 

 

 

 

 

 

EXTernal

“Event In” signal at rear panel BNC

 

 

 

connector, or TRIGger command.

 

 

 

 

 

 

HOLD

Only the TRIGger command will cause

 

 

 

trigger.

 

 

 

 

 

 

IMMediate

The trigger signal is always true (continuous

 

 

 

triggering).

 

 

 

 

• While an instrument which uses the "Event In" signal has EXT set, no other instrument which uses the "Event In" signal may set EXT, or an error 1500 "External trigger source already allocated" will result.

• While TRIG:SOUR is IMM, you need only INITiate the trigger system to start the Pacer.

Related Commands: ABORt, INITiate, *TRG

*RST Condition: TRIG:SOUR IMM

Example Specifying the Trigger Source.

 

TRIG:SOUR HOLD

trigger source is TRIG

 

command

SOUR:PULS:COUN 1E3

output 1000 Pacer pulses

SOUR:PULS:PER .1 S

pulse period set to .1 second

TRIG:DELAY .75 S

start Pacer .75 sec after trigger

INIT

go to Wait For Trigger state

TRIG

trigger the Pacer to output

 

pulses.

:SOURce? TRIGger:SOURce? returns the current trigger source configuration. Response data can be one of; BUS, EXT, HOLD, or IMM. See the TRIG:SOUR command for more response data information.

Example Querying the Trigger Source.

 

TRIG:SOUR HOLD

trigger source is TRIG

 

command

TRIG:SOUR?

ask System Instrument to

 

return trigger source

 

configuration

enter statement

input selection of trigger source

System Instrument Command Reference 7-53

Page 181
Image 181
Agilent Technologies E1300B, E1301B TRIGger SOURce?, Parameter Value Source of Trigger, RST Condition Trigsour IMM