DIAGnostic:INTerrupt:RESPonse?

:INTerrupt:RESPonse?

DIAGnostic:INTerrupt:RESPonse? Returns the interrupt acknowledge

 

response (STATUS/ID word) from the highest priority VXI interrupt line.

Comments

• The value returned is the response from the interrupt acknowledge cycle

 

(STATUS/ID word) of a device interrupting on one of the interrupt lines

 

set up with the DIAG:INT:SET[n] command.

 

• Bits 0 through 7 of the STATUS/ID word are the interrupting device’s

 

logical address. Bits 8 through 15 are Cause/Status bits. Bits 16 through

 

31 (D32 Extension) are not read by the System Instrument.

 

• If only bits 0 through 7 are used by the device (bits 8 - 15 are FF), the

 

logical address can be determined by adding 256 to the value returned by

 

DIAG:INT:RESP?. If bits 0 - 15 are used, the logical address address is

 

determined by adding 65536 to the value returned (if the number

 

returned is negative.

 

 

• Only the interrupt lines previously configured with the

 

DIAG:INT:SET[n] commands generate responses for this command.

 

• If there are interrupts on multiple lines when this command is received,

 

or when the acknowledgement was enabled with DIAG:INT:ACT, the

 

response data returned will be from the line with the highest priority set

 

using the DIAG:INT:PRI [n] command.

 

 

• If interrupt acknowledge has not been enabled with DIAG:INT:ACT,

 

then it will be enabled by DIAG:INT:RESP?. System Instrument

 

execution is halted until the interrupt acknowledgement response is

 

received.

 

 

• DIAG:INT:WAIT? can also be used to wait for the interrupt response.

 

Related Commands: DIAG:INT:ACT, DIAG:INT:SETup[n],

 

DIAG:INT:PRIority[n]

 

Example Setup and wait for VXI interrupt response on line 2.

 

DIAG:INT:PRI2 5

set priority to 5 on line 2

 

DIAG:INT:SETUP2 ON

handle interrupt on line 2

 

.

code which will

 

.

initiate an action

 

.

resulting in an interrupt

 

DIAG:INT:RESP?

read the acknowledge response

7-22 System Instrument Command Reference