Check all settings on a correctly adjusted monitor.

If the mechanical iris is closed more than necessary, the IVS will compensate by increasing the gain and restore the image brightness electronically. This creates additional electronic noise. To avoid this, open the mechanical iris.

• To change the iris, turn the wheel with the iris symbol Õ.

10.3.3Alignment of the image position (X, Y and Rotation) and focus

The position of the image on the CCD Chip and its focus can vary slightly from camera to camera. The video image on the monitor may appear off-centre, not levelled or out of focus.

All adjustments can be made by using a 1,5 mm Allen key, which is delivered with the IVS.

Do not use force!

Image position adjustments in x and y direction can be made with the adjustment screws marked x and y Aphoto.

The screw marked Ò moves the video image horizontally on the monitor. The screw marked Ó moves it vertically.

Ò alignment

Ó alignment

 

Ô alignment

 

 

 

 

 

 

 

 

 

 

F alignment

Note:

Design restrictions imposed by the IVS‘ small

 

size mean that the direction of movement

 

imparted by the x and y adjustment screws

 

might be not precisely in the x and y directions.

Image rotation can be adjusted by the screw marked with the Ô symbol Aphoto.

The image can be focussed using the screw which is marked with the F symbol Aphoto.

Video-Asssist-System