Fluke 435 user manual Volt/Amps/Hertz Measurement Range Resolution Accuracy

Models: 435

1 140
Download 140 pages 63.82 Kb
Page 125
Image 125

 

 

 

Specifications

 

 

 

Electrical Measurements23

ACCURACY, RESOLUTION AND RANGE

 

 

 

 

 

 

 

 

Volt/Amps/Hertz

Measurement Range

Resolution

Accuracy

 

 

 

 

 

 

 

Vrms(AC+DC)

 

 

 

 

 

Fluke 435

1…600 Vrms

0.01 Vrms

± 0.1% of nominal

 

 

 

 

voltage

 

 

600…1000 Vrms

0.01 Vrms

± 0.1%

 

 

Fluke 434

1…1000 Vrms

0.1 Vrms

± 0.5% of nominal

 

 

 

 

voltage

 

 

 

 

 

 

Vpk

1…1400 Vpk

1 V

5% of nominal voltage

 

 

 

 

 

 

 

Voltage Crest Factor

1.0 ... > 2.8

0.01

± 5%

 

 

(CF)

 

 

 

 

 

 

 

 

 

 

 

Arms (AC+DC)

0…20.00 kArms1

0,001…10 Arms1

± 0.5% ± 5 counts3

 

 

Fluke 435

 

Fluke 434

0…20.00 kArms1

0,001…10 Arms1

± 1% ± 5 counts3

 

Fluke 434 with i400s

0…40 / 400 Arms

0.1 and 1 Arms

± 1% ± 5 counts3

 

Fluke 435 with I430flex

30…3000 Arms

1 Arms

± 0.5% ± 20 counts3

 

Apk using 1mV/A scaling

0 - 5500 Apk

1A

± 5%

 

 

 

 

 

 

 

 

A Crest Factor (CF)

1 … 10

0.01

± 5%

 

 

 

 

 

 

 

 

Hz5

 

 

 

 

 

Fluke 435 @ 50Hz

 

 

 

 

 

nominal

42.500 ... 57.500 Hz

0.001 Hz

± 0.01Hz

 

Fluke 435 @ 60Hz

 

 

 

 

 

nominal

51.000 ... 69.000 Hz

0.001 Hz

± 0.01Hz

 

Fluke 434 @ 50Hz

 

 

 

 

 

nominal

42.50 ... 57.50 Hz

0.01 Hz

± 0.01Hz

 

Fluke 434 @ 60Hz

 

 

 

 

 

nominal

51.00 ... 69.00 Hz

0.01 Hz

± 0.01Hz

 

 

 

 

 

 

 

 

 

 

Dips and swells

Measurement Range

Resolution

Accuracy

 

 

 

 

 

 

 

Vrms½ (AC+DC)

 

 

 

 

 

Fluke 435

0.0% ….200% of

0.1Vrms

± 0.2% of nominal

 

 

nominal voltage

 

voltage

 

Fluke 434

0.0% ….200% of

0.1Vrms

± 1% of nominal voltage

 

 

nominal voltage

 

 

 

 

 

 

 

 

 

 

Arms½ (AC+DC)

0 … 20,000 Arms1

 

± 1% ± 10 counts3

 

 

Fluke 435

0,001 Arms…10 Arms

 

Fluke 434

0 … 20,000 Arms1

0,001 Arms…10 Arms

± 2% ± 10 counts3

 

Fluke 434 with i400s

0 … 400 Arms

0.1 Arms and 1 Arms

± 2% ± 10 counts3

 

Fluke 435 with i430flex

30 … 3000 Arms

1 Arms

± 1% ± 20 counts3

 

Threshold levels

Programmable thresholds in percent of nominal voltage

 

 

Event detection based upon ½cycle rms voltages

 

 

 

 

Captures Dips, Swells Interruptions and Rapid Voltage Changes

 

 

 

 

 

 

Duration

hhh,mm,ss,mmm

Half cycle

One cycle

 

 

 

 

 

 

 

23-5

Page 125
Image 125
Fluke 435 Volt/Amps/Hertz Measurement Range Resolution Accuracy, Dips and swells Measurement Range Resolution Accuracy