Fluke 435 user manual Electro Magnetic Compatibility EMC

Models: 435

1 140
Download 140 pages 63.82 Kb
Page 135
Image 135

Specifications23

Electrical Measurements

ELECTRO MAGNETIC COMPATIBILITY (EMC)

Compliance with

Fluke 434/435, including standard accessories, conforms to the EEC directive 2004/108/EC for EMC immunity as defined by EN-61326- 1:2006 , with the addition of the table below

Frequency

No Visible Disturbance

Disturbance < 0.5 %

Disturbance < 10 %

 

 

 

 

80 – 400 MHz

 

All ranges

 

 

 

 

 

 

400

– 600 MHz

 

All other ranges

125 V range

 

 

 

 

 

600

MHz – 1 GHz

 

All ranges

 

 

 

 

 

1.4 – 2 GHz

All ranges

 

 

(3V/m)

 

 

 

 

 

 

 

2 – 2.7 GHz

All ranges

 

 

(1V/m)

 

 

 

 

 

 

 

 

The Analyzer is susceptible for RF fields with a field strength of 10 V/m, between 400 and 600 MHz (Performance criteria B).

1depending clamp scaling, volt scaling 1:1

250Hz/60Hz nominal frequency according to IEC 61000-4-30

3Add clamp accuracy

4The logger and Mains Signaling function are optional for the Fluke 434 and standard available on the Fluke 435

5Measured on reference voltage input A/L1

6Maximum time 9999 hours

7Estimated duration

8Add ±(n-1) x 2.5° for Amp. when using i430flex

23-15

Page 135
Image 135
Fluke 435 user manual Electro Magnetic Compatibility EMC