Fluke 435 user manual Switching from Hold to RUN invokes a menu to select

Models: 435

1 140
Download 140 pages 63.82 Kb
Page 47
Image 47

Harmonics10

Meter screen

 

 

 

 

Switch between HOLD and RUN of screen update.

 

F5

 

 

 

 

 

Switching from HOLD to RUN invokes a menu to select

 

 

 

 

immediate (NOW) or TIMED start time which allows you

 

 

 

 

to define start and duration of the measurement.

 

 

 

 

 

Meter screen

To access the Harmonics Meter screen:

f

F3

The Meter screen display shows 8 measurements per phase. Using the SETUP key and function key F3 - FUNCTION PREF you can choose the screen contents. For detailed information see Chapter 18, FUNCTION PREFerences.

Available function keys:

F1

F3

F4

F5

Selection of harmonics type: Voltage, Current, or Real Power (Watt).

Return to Bar Graph screen.

Access the Trend screen. For description see below.

Switch between HOLD and RUN of screen update. Switching from HOLD to RUN invokes a menu to select immediate (NOW) or TIMED start time which allows you to define start and duration of the measurement.

Trend

To access the Harmonics Trend screen:

g F4

Trend shows how harmonics vary over time: Cursor and Zoom can be used to investigate details. All values in the Meter screen are recorded, but the Trends from each row in the Meter screen are displayed one at a time. Press function key F1 to assign the arrow keys to row selection.

Using the SETUP key and function key F3 - FUNCTION PREF you can choose harmonics display as a percentage of fundamental voltage (%f) or of the total of harmonic voltages (%r, total Vrms). Also the Meter screen contents can be selected in this menu. For detailed information see Chapter 20, FUNCTION PREFerences.