Fluke 435 user manual Screen remains black

Models: 435

1 140
Download 140 pages 63.82 Kb
Page 120
Image 120

Fluke 434/435

Users Manual

Screen remains black.

Make shure that the Analyzer is on: at power-on you should hear a double beep. If the screen remains black, you might have a problem with the screen contrast. Proceed as follows to change Contrast:

Press the SETUP key.

Press Function key F4.

Press the left or right arrow key for five seconds to return to normal display.

Operation time of fully charged battery is too short.

The Battery may be in poor condition. This may improve after a full discharge and full charge cycle as explained in section ‘Keeping the battery in good condition’ in this Chapter.

Printer does not print.

Make shure that the optical interface cable is properly connected between Analyzer and printer.

Make sure that you have selected the correct printer type and printer baudrate. How to proceed is explained in Chapter 21.

If you are using the PAC91 (Print Adapter Cable), make sure that it is turned on and that a fresh battery is installed.

FlukeView does not recognize the Analyzer.

Make sure that the Analyzer is turned on.

Make shure that the optical interface cable is properly connected between Analyzer and PC.

Other PC software does not recognize the Analyzer.

Make sure that the Analyzer is turned on.

Make shure that the optical interface cable is properly connected between Analyzer and PC.

Make sure that the correct COM port has been selected for the PC. If not, change the COM port setting or connect the interface cable to another COM port.

Make sure that baudrate of Analyzer and PC are the same. How to proceed is explained in Chapter 21.

22-4

Page 120
Image 120
Fluke 435 user manual Screen remains black