Fluke 434/435

Users Manual

Mains Signaling

Measurement are based on: either the corresponding 10/12-cycle r.m.s. value

 

interharmonic bin or the rms of the four nearest 10/12-cycle rms value

 

interharmonic bins per IEC 61000-4-30 Limit setup for Monitor mode follows

 

EN50160 “Meistercurve”

 

 

Time Synchronisation

Optional GPS430 timesync module provides time uncertainty ≤ 20 ms or ≤ 16.7

 

ms2 for time tagging of events and time aggregated measurements. When

 

synchoronisation becomes unavailable, time tolerance is ≤ 1-s/24h

 

 

WIRING COMBINATIONS

3Ø WYE

Three phase four wire system WYE

 

 

3Ø DELTA

Three phase three wire system Delta

 

 

1Ø + NEUTRAL

Single phase with neutral

 

 

1Ø SPLIT PHASE

Split phase

 

 

1Ø IT NO NEUTRAL

Single phase system with two phase voltages without neutral

 

 

3Ø IT

Three phase system without neutral WYE

 

 

3Ø HIGH LEG

Four wire three phase Delta system with center tapped high leg

 

 

3Ø OPEN LEG

Open delta three wire system with 2 transformer windings

 

 

2-ELEMENT

Three phase three wire system without current sensor on phase L2 / B (2 Watt

 

meter method)

 

 

2½-ELEMENT

Three phase four wire system without voltage sensor on phase L2 / B

 

 

GENERAL

Case

Design

Rugged, shock proof with integrated protective holster

 

 

Drip and dust proof

IP51 according to IEC60529 when used in tilt stand position

 

 

Shock and Vibration

Shock 30g, Vibration: 3g Sinusoid, Random 0.03g2/Hz according to MIL-PRF-

 

28800F Class 2

 

 

 

 

Display

Bright Full-Color LCD with CCFL backlight, 80cd/m2

Size

115.2 x 86.4 mm

 

 

Resolution

320 x 240 pixels

 

 

Contrast and

User adjustable, temperature compensated

brightness

 

 

 

23-12

Page 132
Image 132
Fluke 435 user manual Case, Display