Display Information 5
Screen information common for all screen types
Screen information common for all screen types
Refer to Figure
A
B
Measuring mode: the active measuring mode is shown in the screen header.
Measuring values: main numerical measuring values. Background colors differ per phase and for voltage or current. If Cursor is on, the values at the Cursor are shown.
C
D
Status indicators. The following symbols may appear on the screen to show the state of Analyzer and measurements:
: Indication that the 150/180 cycle (3 s) aggregation interval (50/60 Hz) is active. With no indication, the aggregation interval is 10/12 cycles (50/60 Hz).
Time that a measurement has been going on. Format: hours, minutes, seconds. When waiting for a timed start, time counts down with prefix
Horizontal ZOOM on.
Measurement may be unstable. E.g. applicable for frequency readout during absence of voltage at reference phase A (L1).
Indicates according to
Recording of measurement data is on.
Phasor rotation / Phase sequence indicator.
Battery/Line power indication. During battery operation the battery charge condition is displayed.
Keyboard locked. Press ENTER 5 seconds to unlock/unlock.
Main area with measuring data: features are explained under 1 ... 5.