Honeywell 2700 manual Calibration Verification, Increasing Load Test

Models: 2700

1 278
Download 278 pages 39.75 Kb
Page 106
Image 106

Step 6. Press the F1 button to activate your selection.

The LCD display shows SCALE CALIBRATE – WAIT SCALE RSP for a few seconds, then it changes to Step 1 – Platter Empty, VOL BTN = Advance. The top LED bar on the scanner flashes once every few seconds.

Step 7. Make sure platter is on correctly and is empty, then press the Sound button.

Step 8. The LCD display leads you through the next step of placing a full weight of 30 lbs. on the scale. The LEDs flash 3 times. Press the Sound button once the scale weight settles on full weight.

Step 9. Remove all the weight once the full capacity is stored. The LED’s flash 5 times. Remove the weight, let the scale settle at zero lbs., then press the Sound button again to accept 0 lbs.

Step 10. The units stores the final weight, and beeps multiple times. The calibration process is complete, and the scale restarts. Begin Calibration Verification, below.

If you have any issues with the calibration, check that the unit has been properly installed and verify that there is no rocking motion when pressing on opposite corners of the scale. If you feel any movement, correct by adjusting the leveling screws that support the scale on the check-stand. When they are correct, lock these down snugly and repeat all the calibration steps.

Once you have completed the scale calibration, the scale must be verified by the appropriate country or state Weights and Measures inspector. For details, contact your local government’s Weights and Measures department.

Calibration Verification

The following tests verify the scale's calibration. The following tests are based on a 2-digit accuracy setting for pounds and a 3- digit accuracy setting for kilograms.

Increasing Load Test

1.Make sure there is no load on the scale platter and verify that the remote display reads:

0.00lbs

or

0.000 kg

2.Place a 5 lb. or a 2.5kg weight on the center of the scale platter and verify the display reads:

5.00lbs

or

2.5kg

3.Place an additional 5 lb. or an additional 2.5kg weight on the center of the scale platter and verify the remote display reads between:

9.99and 10.01 lbs

or

4.995 and 5.005kg

4.Place an additional 10 lb. or 5kg weight on the center of the scale platter and verify the remote display reads between:

19.99and 20.01 lbs

or

9.995 and 10.005kg

5.Place an additional 10 lb. or 5kg weight on the center of the scale platter and verify the remote display reads between:

29.99and 30.01 lbs

or

14.995 and 15.005kg

6 - 10

Page 106
Image 106