10.5.4Analog Outputs Do Not Agree With Front Panel Readings
1.Confirm that the DAC offset
2.Perform a DAC/ADC adjustment and Dark Signal adjustment by following the procedure described in Sections 9.2 and 9.3.
10.5.5Cannot Zero or Cannot Dynamic Zero
1.Check for leaks in the pneumatic system as described in Section 11.3.
2.Confirm that the Zero gas is free of Carbon Monoxide.
3.Check for a dirty particulate filter and replace as necessary as described in Section 11.2.
10.5.6Cannot Span or Cannot Dynamic Span
1.Check for leaks in the pneumatic systems as described in Section 11.3.
2.Check for a dirty particulate filter and replace as necessary as described in Section 11.2.
3.Check for dirty pneumatic system components and clean or replace as necessary.
4.Check for proper adjustment of DAC and ADC electronics by performing the adjustment procedure in Section 9.2.
5.Confirm the sample temperature, sample pressure, and sample flow readings are correct. Check and adjust as required.
10.6Troubleshooting Individual
The following sections provide
10.6.1 Troubleshooting Flow Problems
When troubleshooting flow problems, it is a good idea to first confirm actual flow of the analyzer. If available, use an independent flow meter rotameter or mass flow meter to measure flow(s). Sample flow can be measured at the sample inlet port at the instrument's rear panel. If no independent flow meter is available, placing a finger over an inlet port and feeling for a vacuum will at least give an indication whether flow is present.
If the independent flowmeter shows to be correct, check the pressure/flow sensor assembly as described in Section 10.6.8.