Teledyne 300M 2 Test Function Values, 10-6, Meaning, Acceptable Values, Corrective Action for

Models: 300M

1 123
Download 123 pages 13.29 Kb
Page 100
Image 100

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 10-2: Test Function Values

Test Function

Meaning

Acceptable Values

Corrective Action for

Unacceptable Values

 

 

 

RANGE

The Current Full Scale

Any

None required

 

Range Setting of the

 

 

 

analyzer's analog outputs

 

 

 

 

 

 

CO MEAS

The most recent detector

2500-4800 mV

Check and adjust IR source and

 

reading taken in Measure

 

Sync Demodulator as described

 

mode

 

in Section 10.6.5 and 10.6.6

 

 

 

 

CO REF

The most recent detector

2500 - 4800 mV

Check and adjust IR source and

 

reading taken in

 

Sync Demodulator as described

 

Reference mode

 

in Section 10.6.5 and 10.6.6

 

 

 

 

MR RATIO

The ratio of the CO

1.00 - 1.5

Check CO REF and CO MEAS

 

MEAS value to CO REF

 

values as described above.

 

value

 

 

 

 

 

 

PRES

The absolute pressure of

0" - 1.0" Hg below

Check for pneumatic system

 

the sample gas in the

ambient pressure

problems. See Section 10.6.1.

 

absorption cell

 

Check for pressure transducer

 

 

 

problems. See Section 10.6.5

 

 

 

 

SAMPLE FLOW

Sample flow rate

700 - 900 sec/min

Check for pneumatic system

 

 

 

problems. See Section 10.6.1.

 

 

 

Check for flowmeter problems.

 

 

 

See Section 10.6.5

 

 

 

 

SAMPLE TEMP

The temperature of the

30° - 50°C

See Section 10.6.2

 

sample gas in the

(After warm-up)

 

 

absorption cell

 

 

 

 

 

 

BENCH TEMP

The temperature of the

50°C

See Section 10.6.2

 

Optical Bench

(After warm-up)

 

 

 

 

 

WHEEL TEMP

The temperature of the

68°C

See Section 10.6.2

 

Gas Filter Wheel

(After warm-up)

 

 

 

 

 

BOX TEMP

The temperature inside

up to 10°C above

See Section 10.6.2

 

the analyzer chassis

ambient

 

DCPS

DC Power Supply

2500 ± 200 mV

See Section 10.6.4

 

reference - A composite

 

 

 

of all voltages provided

 

 

 

by the DC Power Supply

 

 

10-6

Page 100
Image 100
Teledyne 300M 2 Test Function Values, 10-6, Meaning, Acceptable Values, Corrective Action for, Unacceptable Values