Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table
# | Signal | Control | Description |
19 | ST_HIGH_RANGE | YES | Status Bit - Autorange High Range |
|
|
| Logic High = M300M in high range |
|
|
| Logic Low = M300M in low range |
|
|
|
|
20 | ST_SYSTEM_OK | YES | Status Bit - System OK |
|
|
| Logic High = No instrument warning present |
|
|
| Logic Low = 1 or more alarm present |
|
|
|
|
21 | ST_BENCH_ALARM | YES | Status Bit - Bench Temperature Alarm |
|
|
| Logic High = Bench Temp out of spec |
|
|
| Logic Low = Bench Temp in spec |
|
|
|
|
22 | ST_SOURCE_ALAR | YES | Status Bit - IR Source Alarm |
| M |
| Logic High = IR Source output too low |
|
|
| |
|
|
| Logic Low = IR Source output normal |
|
|
|
|
23 | ST_WHEEL_ALARM | YES | Status Bit - Wheel Temperature Alarm |
|
|
| Logic High = Wheel Temp out of spec |
|
|
| Logic Low = Wheel Temp in spec |
|
|
|
|
24 | CO_MEASURE | NO | IR detector reading during measure phase. Typically |
|
|
| 2500 - 4500 mV. |
|
|
|
|
25 | CO_REFERENCE | NO | IR detector reading during reference phase. Typically |
|
|
| 2500 - 4500 mV. |
|
|
|
|
26 | SAMPLE_PRESURE | NO | Sample pressure in mV. Typical sea level value = |
|
|
| 4300 mV for 29.9" |
|
|
|
|
27 | VACUUM_PRESSUR | NO | Sample flow in mV |
| E |
|
|
|
|
|
|
28 | SAMPLE_TEMP | NO | Sample temp in mV |
|
|
|
|
29 | BENCH_TEMP | NO | Optical Bench temp. typically 2270 mV for 48°C |
|
|
|
|
30 | WHEEL_TEMP | NO | Filter Wheel temp. typically 4770 mV for 68°C |
|
|
|
|
31 | BOX_TEMP | NO | Internal analyzer temp. in mV |
|
|
|
|