5.2 Diagnostic Tests
The diagnostic tests are used to help diagnose a problem in the analyzer and should only be used by skilled maintenance people since they can potentially interfere with the Carbon Monoxide reading. Table
The TEST button is used to scroll through the test measurements until the one of interest is displayed. To turn the test on, press the OFF/ON button. Viewing test measurements in the diagnostic mode is especially useful for viewing the results of a diagnostic test.
The PREV button goes to the previous diagnostic test. When pressed, the CPU turns the current diagnostic test OFF if it is ON. The NEXT button goes to the next diagnostic test. When pressed, the CPU turns the current diagnostic test OFF if it is ON. The EXIT button exits the diagnostic mode and turns all the diagnostic tests OFF. This ensures that a diagnostic test is not accidentally left ON. A message is also sent to the
Table 5-1: Diagnostic Tests
TEST # | Name | Signal | Nominal Value |
1. | NONE | No output | 0 mV |
|
|
|
|
2. | CO MEASURE | CO detector measure value | 4500 mV |
|
|
|
|
3. | CO | CO detector reference value | 4200 mV |
| REFERENCE |
|
|
4. | SAMP PRESS | Sample pressure | 3650 mV |
|
|
|
|
5. | SAMP FLOW | Sample flowrate | 2000 mV |
|
|
|
|
6. | SAMP TEMP | Sample temperature | 3000 mV |
|
|
|
|
7. | BENCH TEMP | Optical bench temperature | 3900 mV |
|
|
|
|
8. | WHEEL TEMP | Outputs filter wheel temp. | 3900 mV |
|
|
|
|
9. | CHASSIS TEMP | Outputs Chassis temp | 2740 mV |
|
|
|
|
10. | DCPS VOLT | Outputs DC power | 2500 mV |
|
|
|
|
11. | DAS AVERAGE | Current DAS average | Any |
|
|
|
|