Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Examples of possible sequences are as following under any one of three available SEQx.
Example 1: To perform a 15 minute zero check once per day at 10:30 PM, 12/20/93.
1.MODE: ZERO
2.TIMER ENABLED: ON
3.STARTING DATE: 12/20/93
4.STARTING TIME: 22:30
5.DELTA DAYS: 1
6.DELTA TIME: 00:00
7.DURATION: 15
8.CALIBRATE: OFF
Example 2: To perform a 15 min zero span check once per day retarding 15 minutes everyday starting at 11:30 PM, 12/20/93.
1.MODE:
2.TIMER ENABLED: ON
3.STARTING DATE: 12/20/93
4.STARTING TIME: 23:30
5.DELTA DAYS: 0
6.DELTA TIME: 23:45
7.DURATION: 15
8.CALIBRATE: OFF
Example 3: To perform span check once per week starting at 11:30 PM, 12/20/93.
1.MODE: SPAN
2.TIMER ENABLED: ON
3.STARTING DATE: 12/20/93
4.STARTING TIME: 23:30
5.DELTA DAYS: 7
6.DELTA TIME: 00:00
7.DURATION: 15
8.CALIBRATE: OFF