Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
For minimum
Qv × L = 500
Qa × D
Where Qv is the vent flow in cc/min
Qa is the analyzer flow in cc/min
L is the vent line length in inches
D is the ID of the vent line in inches
For Qv of 1000 cc/min
Qa of 800 cc/min
D of .188 inches
L = approximately 72”
8.2 Multi-Point Calibration
The calibration must be carried out:
1.After maintenance
2.Every three months (recommended)
CAUTION
The test gas must be introduced into the analyzer through the sample inlet port. All flow measurement devices must be calibrated against an