Teledyne 300M instruction manual Multi-PointCalibration, Qv × L = Qa × D

Models: 300M

1 123
Download 123 pages 13.29 Kb
Page 81
Image 81
Qv × L = 500

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

For minimum back-diffusion through the vent and for minimum back-pressure in the manifold, the following relationship should be met:

Qv × L = 500

Qa × D

Where Qv is the vent flow in cc/min

Qa is the analyzer flow in cc/min

L is the vent line length in inches

D is the ID of the vent line in inches

For Qv of 1000 cc/min

Qa of 800 cc/min

D of .188 inches

L = approximately 72”

8.2 Multi-Point Calibration

Multi-point calibration requires seven approximately equally spaced calibration points, including zero, using an NIST-traceable CO source.

The calibration must be carried out:

1.After maintenance

2.Every three months (recommended)

CAUTION

The test gas must be introduced into the analyzer through the sample inlet port. All flow measurement devices must be calibrated against an nist-traceable standard such as a bubble-flowmeter that has been calibrated against an nist-traceable volume standard.

8-3

Page 81
Image 81
Teledyne 300M instruction manual Multi-PointCalibration, Qv × L = Qa × D