Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

1.4 Specifications

Ranges

User selectable to any full-scale range from 0 - 30 PPM up to 0 - 3,000 PPM

Zero Noise

< 0.1 PPM (rms)

Span Noise

< 0.5% of reading (rms) or 0.1 PPM (rms)

Lower Detectable Limit

0.2 PPM

Zero Drift (24 hours)

* <0.5 PPM

Zero Drift (7 days)

* <1.0 PPM

Span Drift (7 days)

* 1% of reading

Linearity

1% of Full Scale

Rise/Fall Time (95%)

<60 sec to 95%

Sample Flow Rate

800cc/min. ± 10%

Temperature Range

5-40°C

Humidity Range

0-95%RH, non-condensing

Dust

Sample must be dust free

Temp Coefficient

< 0.5 PPM per °C

Voltage Coefficient

< 0.5 PPM per Volt

Dimensions HxWxD

7" x 17" x 25" (178 mm x 432 mm x 635 mm)

Weight

50 lbs (22.7 kg)

Power

100 50/60 Hz, 115 60 Hz, 220V 50/60 Hz., 240 V 50 Hz. 250 Watts

Power, CE

230 V/50 Hz, 250 Watts

Recorder Outputs

± 100 mV, ± 1 V, ±5 V, ±10 V (Bi-Polar)

Status Output

9 status outputs from opt-isolators

Alarm output

3 opto-isolators to drive TTL relays

* at constant temperature and voltage

1-5