Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
8 CALIBRATION
This section describes a method of performing a
8.1 Required Equipment and Gas Standards
Zero air must be free of CO (less than 0.1 PPM of CO).
CAUTION
Be careful when pulling in outside air particularly if outside humidity and temperature are high. Condensation may result which can lead to unstable operation, or at worst, water contamination in the cell.
Calibration gas concentrations must be generated from an
The materials in the calibration gas delivery system should be stainless steel, TFE and FEP (Teflon). The system must be clean.
The calibration gas delivery system (or manifold) must be properly vented to a suitable vent outside the analyzer area and near the analyzer inlet to avoid imposing a pressure or vacuum at the inlet. The recommended venting method is shown in Figure