Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
3.7 Remote Zero/Span Check or Adjustment (RS-232)
Besides Z/S checking from the front panel, automatic Z/S checking, and remote Z/S checking via the contact closure inputs, the analyzer can also be checked via the
3.8 Power-On Hold Off
Whenever the Model 300M is powered on it will go through a
3.9 Hold Off
Every type of check or adjustment (zero, span, manual, remote, etc.) is followed by a hold off period of from 1 to 20 minutes, during which time the internal data acquisition system (DAS) does not accumulate CO readings into the DAS average. To set the hold off time, press SETUP-