Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 5-2: I/O Signals (Continued)
# | Signal | Control | Description |
8 | WHEEL_HTR | YES | Shows the status of the filter wheel heater. This has |
|
|
| the same function as the LED in the power supply |
|
|
| module. |
|
|
|
|
9 | DARK_CAL | YES | Turns off the detector input to the Sync/Demod board |
|
|
| for electronics calibration. |
|
|
|
|
10 | ELEC_TEST | YES | Activates the Electric Test diagnostic circuitry |
|
|
|
|
11 | BROWNOUT_RESET | YES | Resets the DISP_BROWNOUT circuitry described |
|
|
| above |
|
|
|
|
12 | ST_CONC_ALARM_1 | YES | Status Bit - gas alarm 1 |
|
|
| Logic high = M300M in alarm 1 mode |
|
|
| Logic low = Not in alarm 1 mode |
|
|
|
|
13 | ST_CONC_ALARM_2 | YES | Status Bit - gas alarm 2 |
|
|
| Logic high = M300M in gas alarm 2 |
|
|
| Logic low = Not in alarm 2 mode |
|
|
|
|
14 | ST_FLOW_ALARM | YES | Status Bit - Flow alarm |
|
|
| Logic High = Sample flow out of spec |
|
|
| Logic Low = Flows within spec |
|
|
|
|
15 | ST_TEMP_ALARM | YES | Status Bit - Temperature alarm |
|
|
| Logic High = One or more temps out of spec |
|
|
| Logic Low = Temps within spec |
|
|
|
|
16 | ST_DIAG_MODE | YES | Status Bit - In Diagnostic mode |
|
|
| Logic High = M300M in Diagnostic mode |
|
|
| Logic Low = Not in Diag mode |
|
|
|
|
17 | ST_POWER_OK | YES | Status Bit - Power OK |
|
|
| Logic High = Instrument power is on |
|
|
| Logic Low = Instrument power is off |
|
|
|
|
18 | ST_PRESS_ALARM | YES | Status Bit - Flow alarm |
|
|
| Logic High = Sample pressure out of spec |
|
|
| Logic Low = pressure within spec |
|
|
|
|
(table continued)