Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 5-2: I/O Signals (Continued)

#

Signal

Control

Description

8

WHEEL_HTR

YES

Shows the status of the filter wheel heater. This has

 

 

 

the same function as the LED in the power supply

 

 

 

module.

 

 

 

 

9

DARK_CAL

YES

Turns off the detector input to the Sync/Demod board

 

 

 

for electronics calibration.

 

 

 

 

10

ELEC_TEST

YES

Activates the Electric Test diagnostic circuitry

 

 

 

 

11

BROWNOUT_RESET

YES

Resets the DISP_BROWNOUT circuitry described

 

 

 

above

 

 

 

 

12

ST_CONC_ALARM_1

YES

Status Bit - gas alarm 1

 

 

 

Logic high = M300M in alarm 1 mode

 

 

 

Logic low = Not in alarm 1 mode

 

 

 

 

13

ST_CONC_ALARM_2

YES

Status Bit - gas alarm 2

 

 

 

Logic high = M300M in gas alarm 2

 

 

 

Logic low = Not in alarm 2 mode

 

 

 

 

14

ST_FLOW_ALARM

YES

Status Bit - Flow alarm

 

 

 

Logic High = Sample flow out of spec

 

 

 

Logic Low = Flows within spec

 

 

 

 

15

ST_TEMP_ALARM

YES

Status Bit - Temperature alarm

 

 

 

Logic High = One or more temps out of spec

 

 

 

Logic Low = Temps within spec

 

 

 

 

16

ST_DIAG_MODE

YES

Status Bit - In Diagnostic mode

 

 

 

Logic High = M300M in Diagnostic mode

 

 

 

Logic Low = Not in Diag mode

 

 

 

 

17

ST_POWER_OK

YES

Status Bit - Power OK

 

 

 

Logic High = Instrument power is on

 

 

 

Logic Low = Instrument power is off

 

 

 

 

18

ST_PRESS_ALARM

YES

Status Bit - Flow alarm

 

 

 

Logic High = Sample pressure out of spec

 

 

 

Logic Low = pressure within spec

 

 

 

 

(table continued)