Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
5.3 M300M Internal Variables
The M300M software contains many adjustable parameters. Many of the parameters are set at time of manufacture and do not need to be adjusted for the lifetime of the instrument. It is possible to change these variables either through the
To access the VARS menu press
Table 5-3: M300M Variables
No. | Name | Units | Default | Value | Description | |
Value | Range | |||||
|
|
|
| |||
0 | DAS_HOLD_OFF | MIN | 15 | 0 - 20 | DAS hold off duration after | |
|
|
|
|
| calibration or diagnostic | |
|
|
|
|
|
| |
1 | DYN_ZERO |
| OFF | OFF/ON | Enable to adjust zero | |
|
|
|
|
| calibration through remote | |
|
|
|
|
| contact closure | |
|
|
|
|
|
| |
2 | DYN_SPAN |
| OFF | OFF/ON | Enable to adjust span | |
|
|
|
|
| calibration through remote | |
|
|
|
|
| contact closure | |
|
|
|
|
|
| |
3 | RS232_MODE | Bit Field | 8 | 0 - 99999 | Value is SUM of following | |
|
|
|
|
| decimal numbers: | |
|
|
|
|
| 1=enable quiet mode | |
|
|
|
|
| 2=enable computer mode | |
|
|
|
|
| 4=enable security feature | |
|
|
|
|
| 8=enable front panel | |
|
|
|
|
| menus (API protocol) | |
|
|
|
|
| 16=enable alternate protocol | |
|
|
|
|
| 32=enables | |
|
|
|
|
| support | |
|
|
|
|
| 64=Enable modem | |
|
|
|
|
| 4096=enable command | |
|
|
|
|
| prompt | |
|
|
|
|
|
| |
4 | CLOCK_ADJ | Sec. | 0 | ±60 | ||
|
|
|
|
|
|