Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 1-2: Final Test and Calibration Values

Test Values

 

Observed

Units

Nominal Range

 

Value

 

 

 

 

 

 

 

CO MEAS

 

 

 

mV

2500 - 4700

 

 

 

 

 

 

 

CO REF

 

 

 

mV

2500 - 4700

 

 

 

 

 

 

 

MR RATIO

 

 

 

 

1.1 – 1.22

 

 

 

 

 

 

 

SAMPLE PRESS

 

 

 

in-Hg

27 - 30

 

 

 

 

 

 

 

VACUUM PRESS

 

 

 

in-Hg

10 ± 5

 

 

 

 

 

 

 

SAMPLE FLOW

 

 

 

CC/MIN

700 - 900

 

 

 

 

 

 

 

BENCH TEMP

 

 

 

°C

48 ± 1

 

 

 

 

 

 

 

DC POWER

 

 

 

mV

2500 ± 200

SUPPLY

 

 

 

 

 

SAMPLE TEMP

 

 

 

°C

45-50

 

 

 

 

 

 

 

BOX TEMP

 

 

 

°C

8- 40

 

 

 

 

 

 

 

WHEEL TEMP

 

 

 

°C

68 ± 2

 

 

 

 

 

 

 

 

 

 

 

Diagnostic

 

Dark Ref

 

 

 

mV

125 ± 50

 

 

 

 

 

 

 

Dark Meas

 

 

 

mV

125 ± 50

 

 

 

 

 

 

 

Electrical test

 

 

 

PPM

400 ± 4

 

 

 

 

 

 

 

 

 

Span and CAL Values

 

CO Span Conc

 

 

 

PPM

400

 

 

 

 

 

 

 

CO Slope

 

 

 

 

1.0 ± 0.3

 

 

 

 

 

 

 

CO Offset

 

 

 

 

-0.1 to +0.1

 

 

 

 

 

 

 

Noise at Zero (p-p)

 

 

 

PPM

0.5

 

 

 

 

 

 

 

Noise at Span (p-p)

 

 

 

PPM

<1.0% of reading

 

 

 

 

 

 

 

Factory Installed Options

 

 

Option Installed

Factory Installed Options

 

 

Option Installed

 

 

 

 

 

 

Power Voltage/Frequency

 

 

 

 

 

 

 

 

 

 

Rack Mount, w/ Slides

 

 

 

 

 

 

 

 

 

 

Internal Zero Span Valves

 

 

 

 

 

 

 

 

 

 

Analog Voltage Range

 

 

0 - V

 

 

 

 

 

 

 

PROM #

 

 

Serial #

 

Date

 

 

 

Technician

 

1-16

Page 24
Image 24