Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 1-2: Final Test and Calibration Values
Test Values |
| Observed | Units | Nominal Range | ||
| Value |
| ||||
|
|
|
|
|
| |
CO MEAS |
|
|
| mV | 2500 - 4700 | |
|
|
|
|
|
|
|
CO REF |
|
|
| mV | 2500 - 4700 | |
|
|
|
|
|
|
|
MR RATIO |
|
|
|
| 1.1 – 1.22 | |
|
|
|
|
|
|
|
SAMPLE PRESS |
|
|
| 27 - 30 | ||
|
|
|
|
|
|
|
VACUUM PRESS |
|
|
| 10 ± 5 | ||
|
|
|
|
|
|
|
SAMPLE FLOW |
|
|
| CC/MIN | 700 - 900 | |
|
|
|
|
|
|
|
BENCH TEMP |
|
|
| °C | 48 ± 1 | |
|
|
|
|
|
|
|
DC POWER |
|
|
| mV | 2500 ± 200 | |
SUPPLY |
|
|
|
|
| |
SAMPLE TEMP |
|
|
| °C | ||
|
|
|
|
|
|
|
BOX TEMP |
|
|
| °C | 8- 40 | |
|
|
|
|
|
|
|
WHEEL TEMP |
|
|
| °C | 68 ± 2 | |
|
|
|
|
|
|
|
|
|
|
| Diagnostic |
| |
Dark Ref |
|
|
| mV | 125 ± 50 | |
|
|
|
|
|
|
|
Dark Meas |
|
|
| mV | 125 ± 50 | |
|
|
|
|
|
|
|
Electrical test |
|
|
| PPM | 400 ± 4 | |
|
|
|
|
|
| |
|
|
| Span and CAL Values |
| ||
CO Span Conc |
|
|
| PPM | 400 | |
|
|
|
|
|
|
|
CO Slope |
|
|
|
| 1.0 ± 0.3 | |
|
|
|
|
|
|
|
CO Offset |
|
|
|
| ||
|
|
|
|
|
|
|
Noise at Zero |
|
|
| PPM | 0.5 | |
|
|
|
|
|
|
|
Noise at Span |
|
|
| PPM | <1.0% of reading | |
|
|
|
|
|
|
|
Factory Installed Options |
|
| Option Installed | |||
Factory Installed Options |
|
| Option Installed | |||
|
|
|
|
|
| |
Power Voltage/Frequency |
|
|
|
| ||
|
|
|
|
|
| |
Rack Mount, w/ Slides |
|
|
|
| ||
|
|
|
|
|
| |
Internal Zero Span Valves |
|
|
|
| ||
|
|
|
|
|
| |
Analog Voltage Range |
|
| 0 - V |
| ||
|
|
|
|
|
| |
PROM # |
|
| Serial # |
| ||
Date |
|
|
| Technician |
|