Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
1.6 Electrical and Pneumatic Connections
1.6.1 Electrical Connections
Output #1 Carbon Monoxide concentration - Chart Recorder (REC)
Output #2 Carbon Monoxide concentration - Data Acquisition System (DAS)
Output #3 Test function analog output
Input #4 Zero valve request
Input #5 Span valve request
Input #6 Not Used
There are 2
Outputs 1 and 2 have identical signals and electrical characteristics. Output 3 is the same as 1 and 2 electrically, but has analog TEST function signals routed to it. See Diagnostics in Section 5 for details.
1.6.2 Remote Contact Closures Zero/Span Inputs
Remote contact closures can be used to remotely energize the zero/span valves to do a zero or span check. The external contact closure should be capable of switching 12 VDC at 50 ma.
Refer to Figure
NOTE
Zero or span checks can be performed in any order.
1.6.3 Status Outputs (Optional)
Status outputs report analyzer conditions via contact closures located on the optional