Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 4-1: Setup Functions

Button Sequence

Function

Default

Limits

SETUP-CFG-NEXT

List Software

CO Mach

N/A

 

Configuration

 

 

 

 

 

 

SETUP-ACAL-MODE

Define/Change

Disabled

Zero, Span, Zero-

 

AutoCal Sequences

 

Span

 

 

 

 

SETUP-DAS-EDIT

Define/Change DAS

 

 

 

Data Channels

 

 

 

 

 

 

SETUP-DAS-VIEW

View DAS Data

N/A

N/A

 

 

 

 

SETUP-RNGE-MODE

Set Range Mode

Single

Single, Dual, Auto

 

 

 

 

SETUP-RNGE-SET

Set D/A output range

500 PPM

1-20,000 PPM

 

 

 

 

SETUP-RNGE-UNIT

Set Measurement Units

PPM

PPM. mg/m3, ug/m3

SETUP-PASS

Password Enable

ON

OFF-ON

 

 

 

 

SETUP-CLK-TIME

Set Time-of-Day

00:00

00:00-23:59

 

 

 

 

SETUP-CLK-DATE

Set Current Date

01 JAN 00

31 DEC 99

 

 

 

 

SETUP-MORE-

RS-232 baud rate

2400 baud

300, 1200, 2400,

COMM-BAUD

 

 

4800, 9600, 19.2

 

 

 

 

SETUP-MORE-

Analyzer ID number

0000

0000-9999

COMM-ID

 

 

 

 

 

 

 

SETUP-MORE-

Set Hold-Off Interval

15 min

1 - 60 min

VARS-

 

 

 

DAS_HOLD_OFF

 

 

 

SETUP-MORE-

Enable Remote

OFF

OFF-ON

VARS- DYN_ZERO

Dynamic Zero

 

 

 

Adjustment

 

 

SETUP-MORE-

Enable Remote

OFF

OFF-ON

VARS- DYN_SPAN

Dynamic Span

 

 

 

Adjustment

 

 

 

 

 

 

SETUP-MORE-

Set RS-232 Mode

8

 

VARS- RS232_MODE

 

 

 

 

 

 

 

SETUP-MORE-

Set Clock Adjustment

0

-15 to +15

VARS- CLOCK_ADJ

Rate

 

 

 

 

 

 

4-8

Page 54
Image 54
Teledyne 300M instruction manual 1 Setup Functions, Button Sequence, Default, Limits