10.2.1 Checking the Power Sub-Systems

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

10.2 Troubleshooting Fundamental Analyzer Operation

When the analyzer is turned on, several actions will normally occur which indicate the proper functioning of basic instrument sub-systems. These actions are:

1.The sample pump should start.

2.The green sample light on the front panel should turn on.

3.The display should energize and display a log-on message followed by a standard “Sample" display (See Figure 2-2for illustration of a normal display).

If these actions all occur, it is probable that the analyzer's Power Supplies, CPU, and Display are working properly.

If any of these actions fail to occur, power and/or CPU operation should be checked as follows:

10.2.1 Checking the Power Sub-Systems

WARNING

Hazardous voltages exist within the instrument chassis - use caution.

1.Check incoming line power for proper Voltage and Frequency.

2.Check the Circuit breaker on the analyzer's rear panel.

3.Check the 3-wire safety power-input plug on the analyzer's rear panel.

4.Check for proper internal AC power by confirming that the Red (right-most) LED on the Power Supply Module is lit. If this LED is not lit, replace the fuse at the bottom center of the Power Supply Module.

5.Check for proper DC Voltages by measuring for the following voltages on the V/F Board: +5 V between TP4 and TP5

+15 V between TP1 and TP3 -15 V between TP2 and TP3

If any of these voltages are incorrect, check the DC Power Supply as described in Section 10.6.4.