Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 2-3: Test Measurements
Test Message | Meaning |
TIME=xx:xx:xx | Current |
|
|
RANGE=xxxx PPM | Analog output |
|
|
STABIL=x.xxx PPM | Standard Deviation of CO readings |
|
|
CO MEAS=xxxxx MV | Current V/F measure channel (mV) |
|
|
CO REF=xxxxx MV | Current V/F reference channel (mV) |
|
|
MR RATIO=X.XXX | Ratio of the Reference and Measure values |
|
|
AZERO OFFS=x.xxxxx | Autozero offset |
|
|
PRES=xxx | Absolute sample pressure – inches Hg |
|
|
VAC = xxx | Absolute vacuum pressure – inches Hg |
|
|
SAMPLE FL=xxx CC/M | Sample flow through analyzer (cc/min) |
|
|
SAMPLE TEMP=xxx C | Temperature in the absorption cell |
|
|
BENCH TEP=xxx C | Optical Bench Temperature |
|
|
WHEEL TEP=xxx C | Filter Wheel Temperature |
|
|
BOX TEMP=xxx C | Internal box temperature (degrees C) |
|
|
DCPS=xxxxxx MV | DC power supply (mV) |
|
|
SLOPE=x.xxx | Internal formula - Slope |
|
|
OFFSET=xx.x MV | Internal formula - Offset |
|
|