Teledyne 300M instruction manual 3 Test Measurements, Test Message, Meaning

Models: 300M

1 123
Download 123 pages 13.29 Kb
Page 32
Image 32

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 2-3: Test Measurements

Test Message

Meaning

TIME=xx:xx:xx

Current time-of-day (HH:MM:SS)

 

 

RANGE=xxxx PPM

Analog output full-scale range

 

 

STABIL=x.xxx PPM

Standard Deviation of CO readings

 

 

CO MEAS=xxxxx MV

Current V/F measure channel (mV)

 

 

CO REF=xxxxx MV

Current V/F reference channel (mV)

 

 

MR RATIO=X.XXX

Ratio of the Reference and Measure values

 

 

AZERO OFFS=x.xxxxx

Autozero offset

 

 

PRES=xxx IN-HG-A

Absolute sample pressure – inches Hg

 

 

VAC = xxx IN-HG-A

Absolute vacuum pressure – inches Hg

 

 

SAMPLE FL=xxx CC/M

Sample flow through analyzer (cc/min)

 

 

SAMPLE TEMP=xxx C

Temperature in the absorption cell

 

 

BENCH TEP=xxx C

Optical Bench Temperature

 

 

WHEEL TEP=xxx C

Filter Wheel Temperature

 

 

BOX TEMP=xxx C

Internal box temperature (degrees C)

 

 

DCPS=xxxxxx MV

DC power supply (mV)

 

 

SLOPE=x.xxx

Internal formula - Slope

 

 

OFFSET=xx.x MV

Internal formula - Offset

 

 

2-6

Page 32
Image 32
Teledyne 300M instruction manual 3 Test Measurements, Test Message, Meaning