Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 1-1: Status Outputs
Output# | Pin Pair | Definition | Condition | |
(Low, High) | ||||
|
|
| ||
1 | 1,2 | ALARM 1 | Gas concentration is exceeding the Hi set | |
limit. | ||||
|
|
| ||
|
|
|
| |
2 | 3,4 | ALARM 2 | Gas concentration is exceeding the HiHi set | |
limit. | ||||
|
|
| ||
|
|
|
| |
3 | 5,6 | SYSTEM OK | True if no alarm conditions exist | |
|
|
|
| |
4 | 7,8 | TEMP ALARM | True if any temp warning exists | |
|
|
|
| |
5 | 9, 10 | DIAG MODE | True if in diagnostic mode | |
|
|
|
| |
6 | 11,12 | POWER ON | True if main power is on | |
|
|
|
| |
7 | 13,14 | PRESS ALARM | True on low pressure | |
|
|
|
| |
8 | 15,16 | HIGH RANGE | True if the | |
|
| SELECTED | the analyzer into high range. | |
|
|
|
| |
9 | 17,18 | FLOW ALARM | True if a flow warning exists | |
|
|
|
| |
10 | 19,20 | RESERVED |
| |
|
|
|
| |
11 | 21,22 | SOURCE | True if the analyzer source intensity is out | |
WARNING | of limits. | |||
|
| |||
|
|
|
| |
12 | 23,24 | RESERVED |
| |
|
|
|
|
1.6.4 RS-232
The
1.6.5 Pneumatic System
The Model 300M is equipped with a vacuum pump capable of pulling 800 cc/min across a flow restrictor. This allows a smooth, stable flow of sample through the analyzer.
Sample enters the analyzer through a particulate filter element (47 mm diameter) mounted immediately behind the front panel. The sample then enters directly into the sample cell. Please see Figure