Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 1-1: Status Outputs

Output#

Pin Pair

Definition

Condition

(Low, High)

 

 

 

1

1,2

ALARM 1

Gas concentration is exceeding the Hi set

limit.

 

 

 

 

 

 

 

2

3,4

ALARM 2

Gas concentration is exceeding the HiHi set

limit.

 

 

 

 

 

 

 

3

5,6

SYSTEM OK

True if no alarm conditions exist

 

 

 

 

4

7,8

TEMP ALARM

True if any temp warning exists

 

 

 

 

5

9, 10

DIAG MODE

True if in diagnostic mode

 

 

 

 

6

11,12

POWER ON

True if main power is on

 

 

 

 

7

13,14

PRESS ALARM

True on low pressure

 

 

 

 

8

15,16

HIGH RANGE

True if the auto-range function has switched

 

 

SELECTED

the analyzer into high range.

 

 

 

 

9

17,18

FLOW ALARM

True if a flow warning exists

 

 

 

 

10

19,20

RESERVED

 

 

 

 

 

11

21,22

SOURCE

True if the analyzer source intensity is out

WARNING

of limits.

 

 

 

 

 

 

12

23,24

RESERVED

 

 

 

 

 

1.6.4 RS-232

The RS-232 connection is a male, 9-pin D-sub connector at the location shown in Figure 1-3.See also Appendix A for additional information.

1.6.5 Pneumatic System

The Model 300M is equipped with a vacuum pump capable of pulling 800 cc/min across a flow restrictor. This allows a smooth, stable flow of sample through the analyzer.

Sample enters the analyzer through a particulate filter element (47 mm diameter) mounted immediately behind the front panel. The sample then enters directly into the sample cell. Please see Figure 1-1for a flow diagram and Figure 1-4 for pneumatic connections.