Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Table 2-4: Warning Messages
Warning Message | Meaning |
SYSTEM RESET | Issued whenever analyzer is powered on |
|
|
RAM INITIALIZED | RAM was erased (incl. DAS reports) |
|
|
SOURCE WARNING | IR source < 2500 OR >= 5000 mV |
|
|
AZERO WARNING |
|
|
|
SAMPLE FLOW WARN | Sample flow < 500 cc/m or > 1200 cc/m |
|
|
SAMPLE PRESS WARN | Sample pressure < 15 or > 35 |
|
|
SAMPLE TEMP WARN | Sample temperature < 10°C or > 50°C |
|
|
BOX TEMP WARNING | Box temp. < 12°C or > 48°C |
|
|
BENCH TEMP WARNING | Optical Bench < 43°C or > 53°C |
|
|
WHEEL TEMP WARNING | Filter Wheel < 63°C or > 73°C |
|
|
CANNOT DYN ZERO | CO offset < |
|
|
CANNOT DYN SPAN | CO slope < 0.5 or > 2.0 |
|
|
V/F NOT DETECTED | A/D - I/O card not installed or bad |
|
|
SYNC WARNING | No modulation on detector output |
|
|
DCPS WARNING |
|
|
|
The menu field changes depending on the mode of the analyzer and the buttons that have been pressed. It indicates the current function of each of the 8 push buttons below the display. See Section 2.2.2 for information on using the push buttons.
2.2.2 Programmable Push Buttons
The 8 push buttons below the display are programmable by the CPU and their functions change depending on the mode of the analyzer or the operations being performed. The legend above a button identifies its current function. If there is no legend above a button, it has no function and will be ignored if pressed.