Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
1.7 Operation Verification
The Model 300M Analyzer is now ready for operation.
1.Read Sections 1.3 and all of Section 4 of the manual to understand the analyzer operation.
2.Turn on the power by pressing the on/off switch on the front panel (see Figure
3.After a
4.CO REF, CO MEAS - TEST function values should be between 2500 mV and 4700 mV.
5.Pressure - 29 to 30
6.Sample Flow - 800 cc/min ±20 %.
7.Sample Temp - Ambient temperature (10 - 40°C) ± 10°C.
8.Optical Bench Temp - 48°C ± 1° The computer drives the temp to this setpoint.
9.Filter Wheel Temp - 68°C ± 2° The computer drives the temp to this setpoint.
10.Box Temp - Ambient +10°C.
11.DC Power Supply - 2500 mV ± 200 mV - This is a composite of all of the DC voltages in the instrument. The value is not important but it should be within the range indicated.
12.If the TEST functions are within the limits given above, the instrument should function correctly. If there is a problem, please read the manual and check your setup and test values. The Model 300M is now ready for calibration (see Section 3.1).