Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

If poor correlation exists, check for

1.Flow calculation errors

2.Concentration calculation errors

3.Leaks in manifold

4.Dirt in the manifold

5.Proper manifold venting

6.Zero air system

If none of these help, see the TROUBLESHOOTING SECTION 10.

Record all the analyzer setup data from the display.

Range

DCPS

Box Temp

Wheel Temp

Bench Temp

Sample Temp

Sample Flow

Sample Pressure

Vacuum Pressure

MR ratio (Measure/reference ratio)

CO Reference

CO Measure

These data can be useful in future troubleshooting.