1.5 Installation and Overview

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

1.5 Installation and Overview

The Model 300M is shipped with the following standard equipment:

1.Power cord

2.Operation manual

CAUTION

To avoid personal injury, always use two persons to

lift and carry the Model 300M.

Upon receiving the Model 300M please do the following:

1.Verify no apparent shipping damage. (If damage has occurred please advise shipper first, then Teledyne API.)

2.Before operation it is necessary to remove the shipping hold down screws. Remove the instrument cover, then refer to Figure 1-2for screw location.

3.When installing the Model 300M, allow a minimum of 4 inches (10 cm) of clearance at the back of the instrument and 1 inch (2.5 cm) of clearance on each side for proper ventilation. Also, be sure that the clearance below the chassis is unobstructed by at least the height of the instrument feet.

4.Connect sample inlet line to the sample port on rear panel.

NOTE

See Figure 1-3for rear panel pneumatic connections. Sample gas should only come into contact with PTFE, glass or stainless steel.

Leak check all fittings with soap solution.

CAUTION

Connect the exhaust fitting on the rear panel (see Figure 1-4) to a

suitable vent outside the analyzer area.

5.Connect a recording device to the terminal strip connections on the rear panel (See Figure 1-3).

6.Connect the power cord to an appropriate power outlet (see the serial number tag for correct voltage and frequency).