Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
3.4 Automatic Zero/Span Check
Automatic zero/span checking (Z/S check) must be enabled in the setup mode. The Teledyne API Model 300M Carbon Monoxide Analyzer with Zero/Span Valves option offers capability to check the zero and span point automatically on a timed basis, or through remote
Under the
Mode:
1.DISABLED (Sequence is disabled)
2.ZERO
3.SPAN
4.
To change the setup parameters for a sequence, press the SET button. Pressing the <SET and SET> buttons allows you to scroll through the setup parameters and edit them by pressing the EDIT button. The function of each setup parameter is described below:
Timer Enabled: When set to ON, the sequence will be executed based on the internal timer, as specified in the following parameters. When set to OFF the sequence will be executed only upon an external
Starting Date: The starting date for the sequence is entered in the format of MM/DD/YY, where
MMis the month, DD is the date, and YY is the year. Enter starting date and press ENTR or EXIT to leave the date unchanged.
Starting Time: To set the time of day for the sequence, enter in the format HH:MM, where HH is the hour in