Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 4-1: Setup Functions (Continued)

Button Sequence

Function

Default

Limits

SETUP-MORE-DIAG-

View the state of

N/A

N/A

SIGNAL I/O

internal signals

 

 

 

 

 

 

SETUP-MORE-DIAG-

Generate Analog

N/A

N/A

ANALOG OUTPUT

Output Test Pattern

 

 

 

 

 

 

SETUP-MORE-DIAG-

Calibrate D/A and A/D

N/A

N/A

D/A CALIBRATION

Converters

 

 

 

 

 

 

SETUP-MORE-DIAG-

Generate Electrical

N/A

N/A

ELECTRICAL TEST

Test Output

 

 

 

 

 

 

SETUP-MORE-DIAG-

Adjust Dark Offset

125mV

75 - 175 mV

DARK

 

 

 

CALIBRATION

 

 

 

SETUP-MORE-DIAG-

Calibrate sample

800 cc/m

800 cc/m

FLOW

flowrate

 

 

CALIBRATION

 

 

 

SETUP-MORE-

Select TEST to Analog

None

 

DIAG- TEST CHAN

Output

 

 

OUTPUT

 

 

 

 

 

 

 

SETUP-MORE-DIAG-

Generate RS-232

N/A

N/A

RS-232 OUTPUT

Output Test Pattern

 

 

 

 

 

 

SETUP-MORE-

Enable alarm 1 and set

ON

100 PPM

ALRM-ALM1

limit

 

 

 

 

 

 

SETUP-MORE-

Enable alarm 2 and set

ON

300 PPM

ALRM-ALM2

limit

 

 

 

 

 

 

4-9

Page 55
Image 55