Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Example 4: to perform zero check once per day at 10:30 PM and a span check once per week starting at 11:30 PM, 12/20/93.
1.Select any one of SEQx and program as example 1.
2.Select any other SEQx and program as example.
3.Avoid setting two or more sequences at the same time of the day.
3.5Summary of Front Panel Check and Calibration Controls
The calibration controls are summarized below in terms of the button sequences used to access them.
Table 3-1: Calibration Controls
Button Sequence | Function | Default | Limits |
|
|
|
|
CALZ | Begin zero check | - | - |
|
|
|
|
Adjust CO conc. to zero | - | - | |
|
|
|
|
Exit zero check | - | - | |
|
|
|
|
CALS | Begin span check | - | - |
|
|
|
|
Expected CO span value | 400 PPM | 1 - 20000 PPM | |
|
|
|
|
Adj. CO conc. to span val. | - | - | |
|
|
|
|
Exit span check | - | - | |
|
|
|
|
CAL | Begin | - | - |
|
|
|
|
Adj. CO conc. zero value | - | - | |
|
|
|
|
Expected CO span value | 400 PPM | 1 - 20000 PPM | |
|
|
|
|
Adj. CO conc. to span val. | - | - | |
|
|
|
|
Exit | - | - | |
|
|
|
|
Setup | Disabled |
| |
|
|
|
|
Setup | Disabled |
| |
|
|
|
|
Setup | Disabled |
| |
|
|
|
|