Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
There are two acceptable methods of generating accurate CO concentrations for calibrating the Model 300M.
One method uses a single cylinder of CO
8.2.1Procedure
1.Set the analyzer to the desired range.
2.Set the calibration system to deliver a flow of at least 1000 cc/min. The Model 300M draws approximately 800 cc/min. (See Section 8.1 and Figure
3.
4.Connect the analyzer REC (recorder) terminals to a calibrated
The standard output voltage of
If, in service, data is to be collected from a device (printer) connected to the
5.Set the calibrator to deliver zero air to the manifold.
6.Push “CAL” on the analyzer front panel.
7.Enter password. (If Enabled)
8.Wait 10 minutes for the analyzer to stabilize.
9.Push “ZERO” and “ENTR” on the front panel. The analyzer is now “zeroed.”
10.Push “EXIT”. (Return to sample mode.)
11.Record the DVM reading and the percentage chart reading or the
12.Set the calibrator to produce 75% to 85% of the URL (upper range limit). This will be 400 PPM ±25 PPM on the 500 PPM range.