When the Inactive Pair, Active Pair or specific Wideband service is selected, the Threshold Limits screen will be displayed. This screen displays a list of measurements performed in the selected Autotest and the corresponding pass/fail limits for each measurement.

The limits indicate the values at which the measurement result passes (OK), is marginal (Yield Sign) or is unacceptable (Stop Sign). The lower limit threshold value is shown in the box on the left, the upper limit in the box on the right.

If the test result value passes or is unacceptable, the pass/fail result will correspond to the symbol under the box (either Stop or OK). If the test result value is between the upper or lower limit, the pass/fail result will be marginal (Yield Sign).

Use the [Up] and [Down] keys to highlight the desired test. Use the [Tab] key to move the cursor to the limit value and enter the new value using the blue keys. Press the [Setup] key to restore the factory default values for the selected test. Press the [Enter] key to accept the changes or the [Return] key to ignore all changes and return to the Threshold Setup screen. If invalid limits are entered, a warning screen will be displayed. Press the [Enter] key to return to Edit Limits screen and make appropriate changes to the limits.