2)The temperature is unknown. In this case the screen will show the entered Distance to Strap (or the sum of all sections) in the main result box, and will show the calculated temperature at the bottom. The screen also shows the number of the faulted section at the lower right side of the screen.

If the calculated temperature is much different than the expected temperature, you should suspect that the entered DTS or the entered section information is incorrect.